Fully Automated Clinical Chemistry Analyser & Reagents

  • Analytical methods: end point mono or bichromatic, fixed time, kinetic, differential sample blank.
  • ISE module for: Na+, K+, Li+, Cl- in serum, plasma and diluted urine.
  • 30/43 reagents on board plus diluent.
  • Mono, double reagents method.
  • Reaction temperature: +37°C ± 0.1°C
  • 49 sample/ calibrator/control positions on primary tubes (Ø 12÷13mm/75/100mm)
  • Automatic dilution of the samples.
  • Calibration with linear/non linear curves up to 8 points, auto dilution of the calibration curve
    (linear regression, linear piece-wise, cubic spline, polynomial).
  • Control menu with Levy-Jennings plot.
  • Bar-code reader integrated for samples and reagents (optional).
  • Arm equipped with liquid level detector (capacitive) and shock sensor.
  • High precision micro-metering pump to dispense reagents and samples (no syringe).
  • Reagent volume: 1÷400 microliters.
  • Sample volume: 1÷300 microliters.
  • Typical reaction volume: 200-250 microliters.
  • Wavelengths on board: 8 from 340nm to 700nm plus one free position.
  • Light source: Long life halogen lamp (2000 hr.)
  • Photometric linearity: 0÷3.0 OD.
  • Reading cuvette: 80 on board re-washable made in Bionex©.
  • Cuvette washing: automatic washing station.
  • Water consumption: 2-3 liters per hour.
  • Reagent cooler: 14°C below ambient temp.
  • All in one touch screen computer.
  • L.I.S.: Bi-directional connection.
  • Operative system: Windows© 7.
  • Environment temperature 18-32°C
  • Dimensions: 93x72x60.