Technical Specifications |
|
System |
Random access, batch, STAT |
Throughput |
200/240 tests/hour |
Number of arms |
1/2 |
Number of reagents positions |
30/43 parameters on line plus diluent |
Number of samples / calibrators / controls |
49 |
Analytical methods |
End point, fixed time, kinetic, differential, diff. sample blank, bichromatic |
ISE Module |
yes, Na+, K+, Li+, Cl- (optional) |
Diluter |
1 micro-metering pump |
Reagent volume |
1 ÷ 400 μl |
Sample volume |
1 ÷ 300 μl |
Typical reading volume |
200-250 μl |
Calibration curves |
linear regression, piecewise, cubic spline, polynomial |
Automatic dilution |
Automatic / on request |
Reading cuvettes |
80 automatic re-washable |
Reaction temperature |
+37°C ± 0.1°C |
Wavelengths |
From 340 to 700 nm |
Light source |
Long life halogen lamp (2000 hours life) |
Photometric linearity |
0 ÷ 3.0 OD |
Cuvette washing |
Automatic washing station |
Water consumption |
3 liters / hour |
Reagent cooler |
Up to 14°C below ambient temperature |
Bar-code |
Yes (optional) |
Real-time electronics |
Modular, SMD |
Power supply |
100 ÷240 Vac |
Personal Computer |
External, all in one (Touch screen, PC) |
Peripherals |
Cd rom, HD > 40Gbyte |
Printer |
External laser |
Operational system |
Windows ® |
Interface to LIS host |
Ethernet |
Environment temperature |
18 ÷ 32°C |
Environment RH |
30 ÷ 70 % |
Dimensions |
(WxDxH) 93x72x60 cm |
Weight |
About 80 kg |