MicroELISA Testing FAQ
Frequently asked questions
NEXgen FAQsPersonal LAB FAQsDaily Operation & WorkflowReagents & ConsumablesMaintenance & Troubleshooting
Yes, provided the assay can be programmed on the instrument according to its own IFU. The assay must be validated by the laboratory.
Yes, with a dedicated calibration. The specific plate and assay combination must be validated before routine use.
No. Validated protocols for all EIAgen microELISA kits come pre-loaded on Personal LAB and NEXgen.
12 months from the invoice date against defects in material and workmanship, when the instrument is used according to its manuals and with reagents and consumables supplied by Adaltis.
Up to 5+2: five plates on the working area and two more positions below it. Each plate is processed independently.
Up to 616 primary tubes on sliding trays of 28 tubes each. That is about 600 samples for one assay, or 280 samples for up to 7 different assays at the same time.
Yes. Samples and tips can be loaded continuously while NEXgen is running.
Primary tubes up to 16 mm in diameter and up to 100 mm high.
Yes. NEXgen can combine up to 12 assays in one plate, depending on test compatibility, or run one assay per plate.
Up to 55 reagents and 132 controls, in bottles of up to 60 mL. All resources on board are barcoded and read automatically for positive sample identification (PSID).
Adaltis disposable tips of 200 µL and 1000 µL. 96 reagent tips and up to 9 sample tip racks (864 tips) fit on board, and tips can be loaded continuously during a run.
Yes. The dispensing system detects clots and bubbles during sampling and dispensing, and the disposable tips give zero carry-over.
The incubators sit under the microplates, so the washer and the pipettes reach every plate during incubation. An elevator moves the plates between the incubators, the loading positions and the reader.
Up to 6 buffers in 2.0 L tanks. The 8 L waste tank has remote draining, and gravimetric sensors monitor the level of every tank and the waste.
An LED photometer under the loading area with 8 filter slots (400–700 nm) and a 0–3.3 OD range. It scans each well at 0.2 mm resolution, giving 29 readings across the well, in 5 seconds per plate at one wavelength and 8 seconds at two.
Yes. A dedicated communication module links NEXgen to LIS/LIMS and pre-analytical systems.
Yes. Adaltis service engineers can connect remotely to update calibrations and assay protocols and to help with troubleshooting.
NEXgen measures 130 × 94 × 98 cm (W × D × H) and weighs 220 kg. It runs on 110–230 V AC ±10%, 50–60 Hz, about 450 VA.