NEXgen
Fully Automated 5+2-Microplate ELISA Analyser
The NEW Generation
in Powerful Microplate Automation
Built for laboratories that run many parameters with high sample numbers. Five plates incubate in place on the working area, and two more positions sit below it. Samples and tips are loaded while the analyser runs.
How a run works
From sample tube to result, with no operator steps in between.

Sample station
Up to 96 barcoded tubes load on 16-position strips. The built-in reader identifies every sample before pipetting.

Pipetting
Two independent arms: 1000 µL tips multi-dispense reagents, 200 µL tips handle samples, standards and controls.

Incubation
Two thermally isolated chambers, 30–45 °C, with programmable linear shaking for each plate.

Washing
A 16-channel head (8 dispense, 8 aspirate) with 3 buffers on board, 1–9 cycles and soak up to 3 minutes.

Reading
Single, dual or triple-beam
vertical reading
through 8 interference filters,
0–3.0 OD.
Benefits
616 tubes, continuous loading
12 assays per plate
|
|
|
|
|
|
|
|
|
|
|


Main Features
|
|
|
|
|
|
|
|
|
|
|
|
|
|

Robust and reliable system for a blood bank laboratory
5+2 plates, continuous loading and no plate movement for maximum throughput.
For high-volume, multi-parameter microELISA testing. NEXgen is an open-system, 5+2 microplate ELISA analyser with continuous loading of up to 616 sample tubes. It runs EIAgen kits for HIV, hepatitis and syphilis screening, as well as other validated microplate immunoassays, and connects to LIS/LIMS.
Station by Station
Sample work area
The flexible working area adapts to the needs of today's laboratories and is designed for easy integration with pre-analytical systems as well as LIS/LIMS.
NEXgen can combine different assays in one plate or test one assay per plate, for maximum flexibility in programming and testing.
About 600 samples for one assay, or 280 samples for up to 7 different assays at the same time
Sample sliding trays of 28 primary tubes each, with continuous loading
Microplate platform
The working platform has 5+2 microplate positions and accepts 96-well microplates of all bottom shapes from the main manufacturers.
One shaker (plate rocker) as standard, easily upgraded to up to 4 shaker units
Pipetting station
Two innovative dispensing systems work simultaneously or independently for maximum pipetting efficiency.
Efficient pipetting minimises the drift effect, and standard disposable tips give zero carry-over. The dispensing technology secures precision and accuracy, and identifies and protects against clots and bubbles during dispensing and sampling.
Incubation station
Placing the incubators under the microplates means every assay step runs without moving the plates. The plates stay accessible to the washer and pipettes, even during incubation, which minimises the drift effect.
5 microplates incubated directly on the loading area; 2 more positions under the working area, 1 with shaker
Easy access for both the pipetting system and the wash head
Optional incubator/shaker modules for specific needs, under the sample loading area
An elevator moves microplates to the incubators, loading positions or reader
Temperature: room temperature +7 °C to 50 °C
Reading station
The LED reader, located under the loading area, gives precision, reduced maintenance and a 3.3 OD dynamic range.
Minimal heat output keeps incubation temperatures precise
Long-life LED lamps for minimum maintenance
Large measurement range and excellent reproducibility
High-resolution scanning: each well is measured at 0.2 mm resolution, giving 29 values across the well
Washing station
The washer cleans microplates in the working area while the next microplates are pipetted, so each plate keeps its proper incubation. The independent wash manifold washes each plate directly at the incubation station.
Flexible washing programs adapt processing to virtually any microplate-based assay. Weight sensors report precise volume information for the washing processes on board.
Up to 6 different wash buffers
Waste tank of 8 L with remote draining for easy emptying
Control of the washing flow