| Product Code | |
| Personal LAB | 0-2875 |
| Sample Station | |
| Tubes | Up to 96 positions for tubes 12-13.5 mm diameter, 55 –75 mm height |
| Label | UPC Version A and E (EAN 8 and 13), Code 39, Interleaved 2 and 5, Code 93, Code-bar, Discrete 2 or 5, Code 128, Code 39 FULL ASCII. The scanner will discriminate between the symbols, except Code 39 and Code 39 FULL ASCII. |
| Loading Capacity | 96 samples |
| Sample Identification |
Positive Sample Identification (PSID) by built-in barcode reader and dedicated sample loading facility |
| Sample Loading | Sample strip loading with bar code reading |
| Incubation Station | |
| Incubation Area | Manages 2 microplates (with independent temperature control) |
| Temperature | At least 5°C higher than the room temperature or 30°C ÷ 45°C |
| Stability | ± 1°C |
| Incubation Time | Programmable |
| Warm-Up | 15 minutes |
| Pipetting Station | |
| Working Area | Manages 2 microplates independently |
| XYZ Arm | 2 independent channels using disposable tips for sample, standards, controls and reagents |
| Sample tip type | Adaltis disposable tips |
| Tip size | 200 or 1000 μL |
| Sample pipetting volume | 10-200 μL |
| Reagent pipetting volume | 10-1000 μL |
| Syringe Dilutors | Two dilutors with 1 ml and 5 ml precision micro-pumps |
| Syringe 1 mL resolution | 4000 steps on max. stroke |
| Syringes 5 mL resolution | 2000 steps on max. stroke |
| Replicate Controls / Standards |
Programmable from 1 to 4 |
| Replicate Samples | Programmable from 1 to 8 |
| Precision for Serum Samples |
<5% with 10μL (CV), <3.5% with 25μL (CV) <2.0% with 50μL (CV), <2.0% with 100μL (CV) |
| Precision for Reagents |
<3.5% with 50μL (CV), <2.5% with 100μL (CV), <2.5% with 200μL (CV) |
| Serum Dispensing Time | <15 min. for 88 samples (volume 100μL) |
| Reagent Dispensing Time | <3 min. for 96 wells (volume 100μL) |
| Carryover | Disposable tips: none |
| Washing Station |
|
| Washing Volumes | 100 to 2000 μL |
| Wash Cycles | 1 to 9 |
| Soak time | 1 second to 3 minutes |
| Pressure | Low or normal |
| Photometric Reading | |
| Working Area | Allows for 2 reading stations |
| Reading Range | 0 to 3.0 OD |
| Reading Channels | 16 (2X8) |
| Reading methods | Single, double or double beam with over range filter vertical reading |
| Reading Filters | 8 interference filters |
| Standard Filter Range | 405-450-492-550-620 nm, (other filters optional) |
| Linearity | 1% (0:2 OD) |
| Precision | 1% (0:2 OD) |
| Resolution | 0.001 OD |